Excellent reproducibility is obtainable because dial switches with low contact resistance are used. For example, changes in contact resistance is within +- 1.1m Ohms at 0.1 Ohms setting.
1 m Ohms resolution
Simple, quick dial operation
In-line display for city reading
Ideal for calibration of resistance thermometers and brides
Due to its high accuracy and a dail system, various types of resistance thermometers and bridges can te calibrated accurately and promptly.
Excellent ami shock and -vibration properties
Specifications:
Dielectric Strength 1,000 V AC for one minute between Panel and Circuit
Insulation Resistance More than 500M Ohms at 500V DC between panel and circuit
Dimension 110 x 491 x 140 mm 4 3/8 x 19 3/8 x 5 1/2 in.
Weight 4.8 kg 106 lbs
Resistance Range 0.1m Ohms to 1,111.2100 Ohms (minimum resistance is 0.1 Ohms)
Dial Composition 0.001 Ohms x 10 + 0.01 Ohms x 10 + 0.1 Ohms x 11 + 1 Ohms x 10 + 10 Ohms x 10 + 100 Ohms x 10
Resolution 1m Ohms
Accuracy +-(0.01 Ohms + 2m Ohms) at temperature of 23 +-2 degrees C, humidity of 45 to 85%, and 0.1W Power Application
Max. Allowable Input Power 0.2SW/slep. Within 1W for overall instrument.